# ISO 10993-1:2025 Biological Evaluation of Medical Devices – Edition 6 Update

The ISO 10993-1:2025 standard (Edition 6) for the biological evaluation of medical devices introduces a risk-based, lifecycle approach to assessing biological safety. Published in November 2025, it replaces the prescriptive Table A1 from ISO 10993-1:2018 and aligns closely with ISO 14971 for risk management. Manufacturers now focus on identifying and mitigating actual biological hazards rather than following a checklist of required tests.

## Key Updates in ISO 10993-1:2025

The sixth edition has been completely reorganized to enhance clarity and integration with risk management:

-   **Alignment with ISO 14971:** Biological evaluation is now explicitly integrated into the overall risk management process.
    
-   **Updated terminology:** Terms are clarified to reduce misinterpretation across stakeholders.
    
-   **New guidance:** Includes considerations for exposure duration, materials characterization, and hazard identification.
    
-   **New annexes:** Explain the rationale behind changes and support consistent application by manufacturers and regulators.
    
-   **Transition from Table A1:** The old prescriptive table is replaced by four separate tables emphasizing risk-based evaluation, moving away from a check-box approach.
    

## What Changed from ISO 10993-1:2018?

The most significant shift in ISO 10993-1:2025 is how biological safety testing is approached:

-   **End of the Table A1 approach:** Previously, Table A1 listed required tests by device category, contact type, and exposure duration. This approach is now replaced by a risk-based framework.
    
-   **Lifecycle-based evaluation:** Manufacturers should ask, “What biological safety risks does our device actually present?” instead of “Which tests are mandatory?”
    
-   **Split into four tables:** The new structure better addresses diverse device types and contact scenarios, encouraging comprehensive safety assessments rather than minimum compliance.
    
-   **Focus on scientific justification:** Test selection now depends on material characteristics, intended use, and hazard identification, not just device classification.
    

## Implications for Medical Device Manufacturers

ISO 10993-1:2025 affects regulatory submissions and internal testing strategies:

-   **Enhanced integration with risk management:** Safety evaluation is part of overall ISO 14971 compliance.
    
-   **More flexibility and responsibility:** Sponsors must provide scientific justification for each biological test performed.
    
-   **Potential regulatory adaptation:** EN ISO version is expected soon, which will harmonize requirements across the [EU](https://www.pureglobal.com/markets/european-union).
    
-   **Early planning essential:** Companies should update protocols, dossiers, and testing strategies to align with the risk-based evaluation model.
    

## Action Steps for Compliance

Manufacturers should consider the following actions:

1.  **Review ISO 10993-1:2025 thoroughly** and understand the four new tables.
    
2.  **Map biological hazards to device lifecycle** and exposure scenarios.
    
3.  **Update risk management files** to reflect the integrated approach with ISO 14971.
    
4.  **Justify all selected tests scientifically**, focusing on device-specific risks rather than generic checklists.
    
5.  **Monitor EN ISO publication** to align EU regulatory compliance with the new ISO version.
    