On 26 November 2026, the European Commission published non-binding guidance endorsed by the Medical Device Coordination Group (MDCG) on how notified bodies document reviews under Regulation (EU) 2017/745 (MDR). It addresses two different records. The technical documentation assessment report (TDAR) explains the notified body’s recommendation on certification. The periodic safety update report (PSUR) evaluation records the body’s assessment of the manufacturer’s post-market findings. Commission publication notice.
What the manufacturer should find in the assessment report
The guidance covers reviews of Annex II and III technical documentation under Annex IX sections 4 and 5 and Annex X section 3, including applicable class IIa and IIb sampling under Annex IX sections 2.3 and 3.5 and Annex XI(A) section 10. The notified body may also use it to record reviews of changes during surveillance or recertification. The guidance is specific to the MDR.
For each applicable requirement, the notified body should identify the evidence reviewed and explain how it reached its conclusion. A checklist marked “pass” without acceptance criteria or reviewer reasoning is insufficient. If a deficiency took several rounds to resolve, the report should identify the affected area and reference the review record. Remaining minor nonconformities and follow-up should be visible to the final reviewer.
The TDAR makes a recommendation on the technical file. The final certification decision may also depend on a quality system audit or other assessments. At the end of the process, the notified body provides the manufacturer with the TDAR and clinical evaluation assessment report. MDCG 2026-7, sections 3–6.
When a PSUR evaluation needs an attachment
Section 7 describes the workflow when the relevant EUDAMED functionality is operational. For devices covered by MDR Article 86(2), the manufacturer uploads its PSUR. The notified body records its evaluation and any action taken in the structured fields in every case. If it agrees with the manufacturer’s conclusions, an attached evaluation is optional. If it disagrees, it must also upload its evaluation. The appendix provides a template to identify the device, reporting period and previous version, explain disagreement, and record actions.
Pure Global analysis. A manufacturer preparing its next PSUR can make the comparison easier by identifying what changed in the benefit–risk assessment since the previous report and linking that conclusion to clinical follow-up findings, sales and estimated use. For certification follow-up, use the returned TDAR to identify open issues and the specific file revisions the notified body assessed.