China sets device classes for aesthetic ultrasound products

NMPA Notice 43 links classification to intended use, acoustic output and, for focused ultrasound, focal-plane distance. A device can meet Class III criteria on its intended use alone.

Published on:
November 28, 2026

China’s National Medical Products Administration (NMPA) has issued final principles for classifying products that use ultrasound for medical aesthetics. Notice No. 43 of 2026, dated 19 November and published in Shanghai’s drug regulator policy register on 24 November, applies from publication. Applicants must use the principles to determine whether their product is a medical device and, if so, its class. Notice and official attachment.

Start with the intended use

Section III of the annex covers focused and non-focused ultrasound products that meet the medical device definition and use thermal or mechanical effects for purposes such as fat emulsification, scar repair, skin tightening or pigmentation reduction. By contrast, examples outside device regulation include products used solely for skin cleansing, physical massage or exfoliation that do not meet the device definition. Such products may not claim a medical device intended use.

When do the acoustic thresholds determine the class?

Product or claimClassification under section IV
Focused ultrasound intended to ablate fat by emulsifying or dissolving itClass III, code 01-01.
Focused or non-focused ultrasound with any of the listed scar-repair, lifting, tightening, pore-reduction, skin-elasticity or wrinkle-reduction claimsClass III, code 09-06. One qualifying use is sufficient, even at lower acoustic output.
Other in-scope aesthetic ultrasound devices meeting a Class III parameterClass III, code 09-06, if spatial-peak temporal-average intensity exceeds 100 W/cm² or peak rarefactional pressure exceeds 3 MPa. For focused ultrasound, a focal-plane distance of 1.5 mm or more also qualifies.
Device intended to reduce pigmentation by promoting local blood circulation and accelerating skin metabolismClass II, code 09-06, when intensity is at most 100 W/cm² and pressure is at most 3 MPa. A focused device must also have a focal-plane distance below 1.5 mm.

The Class II criteria require the specified intended use and all applicable parameter limits together. The non-focused criteria omit focal-plane distance. NMPA classification principles, sections III–IV (DOCX).

Registration consequence

For products whose regulatory status or class changes, registrants must follow the subsequent registration arrangements under NMPA Announcement No. 52 of 2026. From 1 December 2029, affected medical aesthetic ultrasound devices lacking a registration certificate for the appropriate class may no longer be manufactured, traded or imported. Notice No. 43, paragraph II.

Pure Global analysis. Review product claims alongside the acoustic specifications for each treatment head. A focused device exactly at 1.5 mm meets the distance criterion for Class III. A focused device below that distance still meets Class III criteria if it claims skin tightening. Record both checks before deciding whether the existing registration covers the product.