EU MDR PSUR: Periodic Safety Update Report

Last reviewed:

Prepare a device-specific PSUR that connects postmarket evidence, benefit-risk conclusions and corrective actions under MDR Article 86.

Regulatory Overview

What is an EU MDR PSUR?

A Periodic Safety Update Report (PSUR) summarizes the results and conclusions of a manufacturer’s post-market surveillance analysis and explains preventive or corrective actions. Under MDR Article 86, it is required for Class IIa, IIb and III devices. It is a lifecycle regulatory record, not a substitute for the PMS plan, complaint investigation or urgent vigilance reporting.

The report should explain whether real-world experience still supports the device’s benefit-risk profile and intended purpose. A collection of complaint counts without denominators, analysis or conclusions is insufficient. MDCG 2022-21 provides the Commission-endorsed practical guidance for MDR PSUR preparation.

PSUR versus PMS Report: which frequency applies?

Regulation and classPeriodic outputMinimum update framework
MDR Class IPMS Report under Article 85Update when necessary
MDR Class IIaPSURWhen necessary and at least every two years
MDR Class IIb and IIIPSURAt least annually
IVDR Class A and BPMS Report under Article 80Update when necessary
IVDR Class C and DPSUR under Article 81At least annually

Apply the IVDR separately; its Class B rule is not the MDR Class IIa rule. Legacy-device applicability and transition conditions also need assessment. Urgent signals, serious incidents and field actions cannot wait for the next periodic cycle.

A practical PSUR evidence map

Report componentEvidence to reconcileQuestion the analysis should answer
Device scope and reporting periodModels, identifiers, versions and grouped-device rationaleWhich devices and time interval do these conclusions cover?
Exposure and useSales, estimated user population and use frequency where practicableWhat denominator makes the event rates interpretable?
Safety/performance signalsComplaints, serious incidents, trends, literature and other PMS sourcesAre rates, severity or patterns changing?
PMCF findingsCompleted activities, limitations and remaining gapsDo clinical data support the claims and residual risks?
Benefit-risk conclusionsRisk-management file, clinical evaluation and state of the artIs the current benefit-risk determination still acceptable?
ActionsCAPA, labeling changes, field actions and future workWhat changed, who owns it, and how will effectiveness be checked?

Use favorable and unfavorable information and explain data limitations, low exposure, missing follow-up and uncertainty. Keep the method consistent across periods or explain why a changed method affects comparison. Grouping devices requires a defensible rationale; a commercial product family is not automatically a suitable reporting group.

Review, availability and EUDAMED

Define data owners, cut-off dates, analysis methods, report author, reviewer and approval authority. Verify the relationship between the reporting interval, device history and the required cadence. Map each conclusion to evidence and each action to a controlled record.

MDR submission/review arrangements differ by device category, including Class III and implantable devices. Confirm the applicable route with the Notified Body and current transitional reporting arrangements. The EUDAMED vigilance/PMS module must not be described as universally mandatory simply because four other modules are in mandatory use.

Connect the PSUR to clinical and quality records

A new safety signal can require changes to risk management, the clinical evaluation, PMCF plan, labeling or design. Record those decisions and their implementation. A PSUR update that leaves a contradictory CER or risk file unchanged is not a complete lifecycle response.

Use the EU MDR service pathway for the wider conformity program and PMS guide for the system generating the data. Pure Global can help define scope, reconcile inputs, prepare the report and coordinate review without presenting a periodic report as an approval of future performance.

How We Can Help

Plan your submission with Pure Global

Plan your device registration, local representation and ongoing compliance with Pure Global.

Pure Global market entry support team

Frequently asked questions

Single Process,
Multiple Markets

When you partner with Pure Global, a single registration process opens doors to multiple countries. Our global subsidiaries make this streamlined path possible.

Let's Talk,
Anywhere You Are.

Whether looking for more information or ready to partner with us, we're here to guide you through every step of the regulatory process.

Contact us