EU MDR CE Marking: Conformity Assessment Guide
Last reviewed:
Plan the MDR conformity assessment, evidence, authorized representation and registration needed for CE marking, with legacy transition conditions assessed separately.
Regulatory background for medical devices in the European Union
In 2021, the EU Medical Devices Regulation (EU MDR 2017/745) became applicable after a four-year transition, replacing the EU Medical Device Directive (MDD) and Active Implantable Medical Device Directive (AIMDD). Devices being placed on the EU market for the first time must now comply with MDR requirements to obtain CE Marking, which is required to distribute medical devices in all EU Member States (though some states have additional requirements). Eligible legacy devices, including certain MDD self-declared devices upclassified under MDR, may continue to be placed on the market only while satisfying all applicable transition conditions according to Regulation (EU) 2023/607, which amended Article 120 of the MDR.
CE Marking Regulatory Roadmap Under the EU MDR
Classification under the MDR determines the conformity assessment procedure for your device. Medical devices are classified into Class I, IIa, IIb, or III based on their risk level, intended use, and the classification rules outlined in Annex VIII of the MDR. Under the MDR, all Class IIa, IIb, and III devices require a conformity assessment conducted by a designated Notified Body (NB) to obtain CE marking. Certain Class I devices also require Notified Body involvement if they are sterile, have a measuring function, or are reusable surgical instruments (Class Is, Class Im, Class Ir).
EU MDR Conformity Assessment Procedures for Medical Devices
All medical devices, regardless of classification, must meet the following key requirements to obtain CE Marking under the EU MDR and maintain regulatory compliance throughout the life cycle of the device:
fulfill the General Safety and Performance Requirements (GSPRs) outlined in Annex I of the MDR.
compile and maintain Technical Documentation in accordance with Annexes II and III that demonstrates compliance with MDR requirements.
implement a Quality Management System (QMS) that aligns with MDR requirements, typically conforming to ISO 13485:2016.
assign a Unique Device Identifier (UDI) to enable traceability throughout the supply chain.
non-EU manufacturers must appoint an EU-based Authorized Representative (symbol EU REP) to represent them to regulatory authorities.
Manufacturers of Class I medical devices (excluding sterile, measuring, or reusable surgical instruments) may self-declare conformity and affix the CE Mark without the involvement of a Notified Body. However, all Class Is (sterile), Im (measuring), Ir (reusable surgical), IIa, IIb, and III devices require a Notified Body audit.
The scope and depth of the NB conformity assessment increases with the risk classification. For example, manufacturers of high-risk devices requiring pre-market clinical investigations can expect the NB to scrutinize their clinical data and verify that the investigation was conducted to ensure clinical compliance with Articles 62-82 of the MDR. Risk management files, QMS documentation, supply chain, and Clinical Evaluation Reports will be thoroughly examined. However, for low-risk devices, such as a Class I measuring device, the NB audit will be limited to conformity procedures relevant to the measuring function.
Where Notified Body involvement is required and the assessment is successfully completed, the body issues the applicable conformity assessment certificate. The manufacturer completes and signs the EU Declaration of Conformity before affixing the CE marking and must complete applicable registration before market placement, including the need to register the device in EUDAMED and/or with the relevant Competent Authority.
CE marking release checklist
Before release, reconcile the GSPR evidence map, risk file, final clinical evaluation, approved labels and the actual certified device range. Confirm the certificate's scope, validity and any conditions. A sterile, measuring or reusable-surgical Class I assessment is limited to the relevant aspects; it is not the same review as Class IIa–III conformity assessment.
Maintain PMS and PSUR arrangements after marking. For legacy products, retain evidence that transition conditions were met on time; a final end date alone does not establish eligibility.
Official framework and connected work
Use the current MDR, IVDR and Commission guidance for the applicable legal text and interpretation. Start from the EU market-entry hub for the overall pathway; this guide addresses the specific task within that program.
For regulatory terminology and the authority framework, see the EU MDR.
Start selling your medical device in the European Union
Plan your device registration, local representation and ongoing compliance with Pure Global.

Frequently asked questions
Ordinary Class I devices can follow self-declaration, while sterile, measuring and reusable surgical Class I devices require Notified Body involvement limited to the relevant aspects. Classification, technical documentation and manufacturer duties still apply.
No. The manufacturer must meet the applicable MDR requirements, complete the conformity assessment, maintain device-specific evidence and issue the EU Declaration of Conformity. A quality-system certificate alone does not authorize the device.
Eligible legacy devices must satisfy Article 120 conditions, including no significant design or intended-purpose changes and no unacceptable risk. The MDR QMS and formal Notified Body application milestones were 26 May 2024, with the written agreement due by 26 September 2024. Eligible higher-risk devices have an end date of 31 December 2027 and the specified other groups 31 December 2028; the Class III custom-made implantable transition ended on 26 May 2026. Check the exact device scope and every condition rather than treating an end date as a blanket extension.